Test voltage | 125/250/500/1000V |
---|---|
Measuring range (Auto range) |
4.000/40.00/200.0MΩ(125V) 4.000/40.00/400.0/2000MΩ(250/500/1000V) |
Mid-scale value | 5MΩ (125/250V) 50MΩ (500V/1000V) |
First effective measuring range | 0.2 - 20MΩ (125V) 0.2 - 40MΩ (250V) 0.2 - 200MΩ (500V) 0.2 - 1000MΩ (1000V) |
Accuracy | ±2%rdg±6dgt |
Second effective measuring ranges |
0.110 - 0.199MΩ 20.01 - 200.0MΩ (125V) |
Accuracy | ±5%rdg±6dgt |
Rated current | DC 1 - 1.2mA |
Output short circuit current | 1.5mA max |
Ω / Continuty
Auto range | 40.00/400.0Ω |
---|---|
Accuracy | ±2%rdg±8dgt |
Output voltage on open circuit | 5V±20% |
Output short circuit current | DC 220±20mA |
Fuse | Quick acting ceramic fuse 0.5A/600V (Ø6.35×32mm) |
AC Voltage
AC Voltage range | AC 20 - 600V (50/60Hz) DC -20 - -600V/+20 - +600V |
---|---|
AC Voltage | ±3%rdg±6dgt |
General
Applicable standards | IEC 61010-1 CAT III 600V, IEC 61557-1,2,4, IEC 61326-1 (EMC), IEC 60529 (IP40) |
---|---|
Power source | R6 × 6 or LR6 × 6 |
Dimensions / Weight | 105(L) × 158(W) × 70(D)mm / 600g approx. |
Included Accessories | 7150A (Remote control switch set), 9121 (Shoulder strap), R6 (AA) × 6, Instruction Manual |
Optional Accessories | 7115 (Extension Probe), 8016 (Hook type prod), 8923 (Fuse [0.5A/600V]), 9089 (Carrying Case) |